Description of Calibration Log screen
Past calibration history is displayed.
Sensor input value, concentration value, and
the date when zero/span calibration is performed
are logged. The 10 newest calibration data are
logged by each component.
Move the cursor to Clear Calibration Log and
ENT
press the
key, and the calibration log is
cleared completely.
Z1 : Zero calibration (Z) of Range 1
S1 : Span calibration (S) of Range 1
Cnt : Value of measuring detector at the time
of calibration
Con : Concentration value displayed before
calibration