Standard Functions; Optional Functions - Fuji Electric ZPB Insruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

2. Standard Functions

Output signal holding:
Output signals are held unchanged during
manual and auto calibrations by activation
of holding (turning "ON" its setting).
The values held are those just before
start calibration mode or setting value.
Usage is selectable.
Indication of instantaneous values will
not be held.
Switch ranges: The switch ranges function is available in
manual, auto, and remote modes. Only
preset switch method is effective.
second range when the measured value
(Option)
signal when remote range switch input
is received.
When the contact input terminals for
each component are input voltage, the
to the second range when the terminals
are open.
* These switch range value are settable between
range values).

3. Optional Functions

Remote output holding:
Output signal is held at the last value
or preset value by voltage input to the
remote output holding input terminals.
Holding is maintained while is voltage
input to the terminals. Indication of in-
stantaneous values are not held.
by a contact position.
The contact output terminals close for
is selected, and open when the second
range is selected.
Auto calibration:
at the preset cycle.
When a standard gas cylinder for calibra-
tion and a solenoid valve for opening/
ternally by the customer, calibration will be
carried out with the solenoid valve drive
contacts for zero calibration and each
span calibration turned on/off sequentially
at the set auto calibration timing.
INZ-TN2ZPB-E
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day).
in auto calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
Auto calibration remote start:
auto calibration remote start input terminal
after short circuiting for 1.5 sec or longer.
open.
Auto zero calibration:
odically at the preset cycle.
calibration" cycle.
When zero calibration gas and solenoid
valve for opening/closing the calibration
the customer, zero calibration will be car-
ried out at the set auto zero calibration
timing.
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day)
zero calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
High/low limit alarm:
measurement value reaches the preset
high or low limit alarm value.
Contacts close when the instantaneous
value of each channel exceeds the high
alarm limit value or falls below the low
alarm limit value.
Instrument error contact output:
Contacts turn on at occurrence of analyzer
error No. 1, 2, 3 or 10.
Calibration error contact output:
Contacts turn on at occurrence of manual
or auto calibration error (any of errors No.
4 to 9).
Auto calibration status contact outputs:
Contacts turn on during auto calibration.
O
correction:
Correction of measured NO, SO
-
2
CO gas concentrations into values at
reference O
77
and
2
concentration.
2
-