Fuji Electric ZPB Insruction Manual page 74

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

6.8.2 Manual span calibration
SPAN
ENT
plied.
CAUTION
For the Ch (components) in which "both" is
set in the calibration range setting (Refer to
Section 6.2.3) - span calibration is completed
for both ranges.
ENT
Note: For the Ch (component) for which
"AR" is selected in Section 6.1.1
"Setting of range switch mode",
the cursor automatically moves to
the range selected in "Setting of
auto calibration component/range"
(Section 6.2.4), and calibration is
carried out within that range.
To close "Span Calibration"
To close the "Span Calibration" or cancel
this mode midway, press the
A previous screen will return.
measure-
-
sensor.
SPAN Cal.
-
SPAN Cal.
key and press
-
SPAN Cal.
ESC
key.
64
X
Measurement Mode Screen
SPAN
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-50.00
Ch4
Range 2 0-500.0
CO
Range 1 0-10.00
Ch5
Range 2 0-25.00
O
2
(
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch4
Range 1 0-50.00
Range 2 0-500.0
CO
Range 1 0-10.00
Ch5
Range 2 0-25.00
O
2
ENT
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-50.00
Ch4
Range 2 0-500.0
CO
Range 1 0-10.00
Ch5
Range 2 0-25.00
O
2
ENT
To Measurement screen after
executing Manual Span Calibration
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.00
vol%
ppm
0.0
ppm
vol%
vol%
20.09
ENT
)
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.00
vol%
0.0
ppm
ppm
vol%
vol%
20.09
ppm
0.0
ppm
ppm
0.9
ppm
vol%
0.34
vol%
ppm
ppm
vol%
vol%
20.09
INZ-TN2ZPB-E