Description of output adjustment screen
Analog output adjustment screen.
Connect the digital multi meter to the output
terminal corresponding to the number of OUT
to be adjusted, and adjust the value so that 4mA
or 0V is output at zero and 20mA or 1V is
output at span.
ENT
key.
ENT
key.
Maintenance
Mode
Output Adj.
OUT
Zero
1
00600 03700
2
00600 03700
3
00600 03700
4
00600 03700
5
00600 03700
6
00600 03700
Maintenance
Mode
Output Adj.
OUT
Zero
1
00600
2
00600 03700
3
00600 03700
4
00600 03700
5
00600 03700
6
00600 03700
56
Adjust OUTPUT
ZERO and SPAN
Span
OUT
Zero
Span
7
00600 03700
8
00600 03700
9
00600 03700
10
00600 03700
11
00600 03700
12
00600 03700
ENT
Zero / Span adjustment
Span
OUT
Zero
Span
03700
7
00600 03700
8
00600 03700
9
00600 03700
10
00600 03700
11
00600 03700
12
00600 03700