Fuji Electric ZPB Insruction Manual page 68

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

ENT
key.
Settable range
The value for range 1 and range 2 must fall within
the range from the MIN and the MAX range
(including the MIN and the MAX range), and
at the same time range 1 must be smaller than
range 2.
The number of ranges is 1 or 2.
key.
CAUTION
Be sure to perform zero / span calibration when the range setting is changed.
Otherwise, the measurement value may not be output properly.
or
ENT
key.
ENT
key.
Maintenance
Mode
setting
Password set
O
ref. Value 12% O
2
Station No. 01
Range setting
-
Maintenance
Mode
Range set
Ch1
NOx
Ch2
SO
Ch3
CO
Ch4
CO
Ch5
O
Maintenance
Mode
Range set
Ch1 NOx
MIN range
Range 1
Range 2
MAX range
Range num.
Maintenance
Mode
Range set
Ch1 NOx
-
MIN range
Range 1
Range 2
MAX range
Range num.
ENT
58
Select an item
2465
limit 20% O
2
2
ENT
Select Ch No.
2
2
2
ENT
Select range or
range num.
50.00
ppm
100.0
ppm
300.0
ppm
500.0
ppm
2
ENT
Set range
50.00
ppm
100.0
ppm
300.0
ppm
500.0
ppm
2