6-1-3 Wiring to Bridge Box

How to configure a variety of strain gage bridges is described below.
Gage Method
Quarter bridge system
(2-wire)
Quarter bridge system
(3-wire)
Half bridge system
Half bridge system
(3-wire active-active)
Full bridge system
NOTE
When wiring strain gage (s) to the bridge box, soldering is recommended to reduce contact
resistance.
Use shortest possible leadwires for connecting the strain gage and the bridge box. Leadwires
shorter than 1 m are ideal.
Use a shielded cable for dynamic strain measurement. It is advantageous to reduce noise.
For other details, refer to the bridge box's Instruction Manual.

6-1-4 Shield Conversion Cable N-117

When using a bridge box that can input multiple channels
with DPM-911/912/913, please use shield conversion cable
N-117 (optional).
It can be isolated for each channels.
Bridge Circuit
33
Wiring to the bridge box