5-12 APPLICATION OF CENTRALIZED CONNECTOR

The DPM-900/950 series provides a centralized connector on its rear panel. Generally, the connector is
used to connect with an AC power supply via the accessory power cable. The connector also provides a pin
that connects with a DC power supply. The connector also enables connection of input/output signals.
Names and functions of these signals are shown in the following table.
For details of external control signals (with REM on the signal name) in the table, read "5-13 REMOTE
CONTROL."
Pin No.
Signal Name
1
+12 V
2
REM BAL
3
OSC
4
N.C
3
5
COM
6
N.C
3
7
AC POWER INPUT
3
8
N.C
9
Bridge status
10
REM KEYLOCK
11
GND
12
N.C
13
N.C
14
0 V
15
OUTPUT A
16
COM
17
REM +CAL
18
REM -CAL
3
19
N.C
20
100 VAC
Connector plug: HIROSE ELECTRONIC S-1620A-C
1
Use a heat-resisting vinyl cable of over 0.3 mm
2
It is a signal with Pin #5 COM as common electric potential. (COM and GND have same electric
potential)
3
N.C: Non connection

5-13 REMOTE CONTROL

The following functions can be controlled remotely.
Execution of balancing
Outputting CAL
Execution and cancellation of key-lock
Remote control has the same significance as manual operation of the front panel switches that control the
DPM-900/950 series functions. The priority is given to the first operated one.
Signal type
Power supply
2
Remote input
Input/output
Common electric
potential
Power supply
Output
2
Remote input
Grounding terminal
Power supply
2
Output
Common electric
potential
2
Remote input
2
Remote input
Power supply
Connects
1
with the + terminal of a DC power
supply.
Valid with an external balancing signal "L."
Connects with input synchronous signal.
Common electric potential of the remote signal,
bridge status, OUTPUT A
Also can be used as OCS – signal.
Connects with an AC power supply.
Outputs open collector
check which has proven N.G.
Valid with an external key-lock signal "L."
Connects with the case and COM terminals
internally.
Connects with the minus terminal of a DC power
1
supply.
Connects with OUTPUT A connector internally.
Common electric potential of the remote signal,
bridge status, OUTPUT A
Also can be used as OCS – signal.
Valid with an "L" signal that commands execution
of external +CAL input.
Valid with an "L" signal that commands execution
of external -CAL input.
Connects with an AC power supply.
2
in diameter for wiring.
REM BAL signal
REM +CAL, REM –CAL signals
REM KEYLOCK signal
27
Function and destination
*12
"L" against a bridge