S/N Ratio
when 500 m/m is input,
outputs 10.00 V.
when 1000 m/m is input,
outputs 10.00 V.
Output
Stability
Zero - Temperature
Zero - Time
Zero – Power Fluctuation
Sensitivity - Temperature
Sensitivity - Time
Sensitivity – Power Fluctuation
Withstand Voltage
Output Voltage Display
Excessive Input Display
Check Function
Key-lock Function
Remote Function
Synchronization Method
TEDS
Actual Load Calibration
Vibration Resistance
Impact Resistance
Operating Temperature &
Humidity Range
Storage Temperature Range
Power Supply
Dimension
Weight
DPM-911/951
911 54 dBp-p or more
(RTI: Within 2 m/mp-p)
951 53 dBp-p or more
(RTI: Within 2.3 m/mp-p)
911 60 dBp-p or more
951 58 dBp-p or more
[Common Condition] Excitation Voltage: 2 Vrms,
Bridge Resistance: 120 ohm, LPF = FLAT
OUTPUT A: ±10 V (Load resistance 5 kohm or more)
OUTPUT B: ±10 V (Load resistance 5 kohm or more)
±0.1 m/m/ºC
±0.5 m/m/24h
±0.05% FS/±10% fluctuation
±0.05%/ºC
±0.3%/24 h
±0.05%/±10% fluctuation
Between measuring bridge – case : 1000 VAC for 1 minutes
Between AC power – case : 1000 VAC for 1 minutes
1
4
/
-digit digital display (7-segment LED)
2
11-segment LED bar graph meter
Output Voltage Display flickers. (4
Setting methods other than POWER switch become invalid
(However, the setting values of the CAL and VOLTAGE OUT switches
Capable of controlling the following functions
Addition of the calibration signal (CAL)
Automatically determines internal (INT) or external (EXT)
Read the TEDS information of the TEDS-compatible sensor and set the
rated output to the output voltage of the VOLTAGE OUT switch
(Condition: Within the setting range of the sensitivity adjuster)
Set the actual load to the output voltage of the VOLTAGE OUT switch
(Condition: Within the setting range of the sensitivity adjuster)
X, Y, and Z directions: 12 cycles with 10 minutes/cycle
X, Y, and Z directions: each 3 times
-10 to 50ºC, 20 to 85 %RH (noncondensing)
90 to 110 VAC (approximately 10 VA: 100 VAC)
10.5 to 15 VDC (approximately 0.6 A: 12 VDC)
The following products are also available upon request.
A115: 108 to 132VAC
A200: 180 to 220VAC
A230: 207 to 253VAC
49 (W) mm× 128.5 (H)mm × 262.5 (D) mm (Protruded portion not included)
40
DPM-912/952
912 52 dBp-p or more
(RTI: Within 2.5 m/mp-p)
952 47 dBp-p or more
(RTI: Within 4.5 m/mp-p)
912 57 dBp-p or more
952 53 dBp-p or more
1
/
-digit digital display only)
2
Bridge Check
can be changed.)
Balance adjustment (BAL)
Key-lock
and manual setting
2
5 to 200 Hz, 29.4 m/s
15 G, 11 msec or less
-30 to 70ºC
Approximately 1.2 kg
DPM-913
48 dBp-p or more
(RTI: Within 3.9 m/mp-p)
53 dBp-p or more
±0.2 m/m/ºC
±1.0 m/m/24h
±0.05% FS/±10% fluctuation
±0.05%/ºC
±0.3%/24 h
±0.05%/±10% fluctuation
(3G)