4. BASIC OPERATION

This chapter describes the basic operation of the DPM-900/950 series.
The instructions given in this chapter is related to the usage of a single unit.
MEMO
For operation of multiple channels using the exclusive carrying case, read "6-5 EXCLUSIVE
CARRYING CASE."

4-1 CONNECTION OF INPUT

Connect a strain gage (via a bridge box) or a strain
gage transducer to the DPM-900/950 series' INPUT
connector.
Configure a Wheatstone Bridge with the strain gage(s)
and the bridge box.
If one end of the connection cable of a strain gage
transducer is formed by a NDIS connector, insert it
into the INPUT connector directly as shown at right.
If the connection cable of a strain gage transducer comes without the NDIS connector on its end or if a
strain gage is to be connected directly, either of the following 2 methods can be used.
1) When using the bridge box for connection
Read "6-1 BRIDGE BOX."
2) When connecting the strain gage to the plug in advance
A connection method is shown in the right. The plug is NDIS
connector PRC03-12A10-7M10 (TAJIMI).
For connection of a strain gage, a bridge box of full bridge
method
*5
can be conveniently used.
MEMO
For the E terminal, read "9-5 E TERMINAL OF
INPUT CONNECTOR."
Wire colors described in figures are typical colors of
KYOWA strain gage transducer.
NOTE
Do not connect a remote sensing sensor to the DPM-900/950 series. The F terminal and G terminal are
input terminals dedicated for the TEDS.
When connecting the cable, hold the plug and
push it in the jack.
When disconnecting, pull the plug toward you.
Transducer Case
TEDS
12
Bridge Box
Strain Gage
Transducer
Shield
Green
White
Black
(+)
(-)
or
NDIS Connector
G
Red
Yellow
Blue