Maintenance
Servicing Plan
Table 8-1.
Maintenance procedures
Item
Forepump Sogevac
Chiller
8-6
iCAP Q Operating Manual (P/N 1288090, Revision B)
Procedure
Check oil level
Check gas ballast valve
Replace exhaust filter
Check anti-suck back valve
Clean fan cover
1st oil change
Replace PFPE oil
Replace fluid bag
Replace condenser filter
Clean reservoir
Frequency
Daily
Monthly
Annually
Annually
Annually
Depending on operation
condition
Annually or depending
on operation condition
Depending on operation
condition
Depending on operation
condition
Depending on operation
condition
Procedure location
see manufacturer's
manual
see manufacturer's
manual