System Description
Gas Supply
Argon
3-18
iCAP Q Operating Manual (P/N 1288090, Revision B)
For some applications, the use of additional argon gas is beneficial. The
gas requirements are detailed in
Table 3-7.
Additional gas requirements for argon
Gas Requirements
Purity
Maximum water content
Supply rate
Pressure regulator output
Table
3-7.
Specification
99.996%
< 5 ppmv
max. 1 L/min
0.6 MPa (6 bar)