Thermo Scientific iCAP Q Operating Manual page 58

Table of Contents

Advertisement

System Setup
Placing the System
Minimum Clearance
5-6
iCAP Q Operating Manual (P/N 1288090, Revision B)
Allow at least 70 to 100 cm of clear space left of the system for proper
air circulation and for clearance of the gas lines, electrical connections,
and exhaust lines and vacuum hose of the fore vacuum pump.
On the left side panel, free access to the mains switch and circuit breaker
is needed to allow shutting off the instrument in case of an emergency at
all times.
It is recommended to leave sufficient space on the right side of the
iCAP Q system for the connections to the computer and peripheral
devices.
NOTICE Avoid blocking the ventilation slots at the side of the
instrument. Items may fall behind the instrument, inhibit airflow, and
cause the system to overheat.
Thermo Fisher Scientific