Thermo Scientific iCAP Q Operating Manual page 114

Table of Contents

Advertisement

Maintenance
Maintaining the Interface
Cleaning the Extraction Lens
CAUTION
!
8-24
iCAP Q Operating Manual (P/N 1288090, Revision B)
Figure 8-17. Pull out the extraction lens
5. Dismount the extraction lens assembly.
Figure 8-18. Extraction lens
6. Check the extraction lens parts for damage or dirt:
Either, clean the extraction lens, see
Lens"
on
page
Or, replace the extraction lens.
Corrosive Acids. When handling the corrosive acids required to
perform the cleaning procedures described below, take the following
precautions.
• Wear protective clothing.
• Use acid resistant laboratory gloves and glasses.
• Make yourself aware of all safety procedures in case of spillage or
exposure to skin or eyes before commencing the cleaning.
The acid cleaning should take place in a fume hood.
"Cleaning the Extraction
8-24,
Thermo Fisher Scientific