Maintenance
Maintaining the Interface
8-18
iCAP Q Operating Manual (P/N 1288090, Revision B)
Figure 8-8. Cone tool at cone housing
c. Slightly turn counterclockwise to unscrew the sample cone and
the sample cone holder.
Figure 8-9. Sample cone and holder in cone tool
d. Remove the sample cone and the cone holder from the tool.
The sample cone can now be cleaned.