Thermo Scientific iCAP Q Operating Manual page 112

Table of Contents

Advertisement

Maintenance
Maintaining the Interface
8-22
iCAP Q Operating Manual (P/N 1288090, Revision B)
Figure 8-14. Insert tool and skimmer cone insert
2. Turn the tool to remove the insert from the skimmer cone.
3. Carefully remove the insert.
4. Clean the insert. Proceed as described in
Cone and the Skimmer Cone"
of nitric acid.
5. After the cleaning procedure is completed, position the insert at the
tool again, see
Figure
Figure 8-15. Skimmer cone insert and tool.
NOTICE Make sure to mount the insert with the correct side facing the
tool, see
Figure
8-15.
"Cleaning the Sample
on
page 8-19
8-15.
using a 10% solution
Thermo Fisher Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Icap qaIcap qcIcap qs

Table of Contents