8. Technical Specifications

Unless otherwise specified, the technical specifications applied are for the
oscilloscope only, and Probes attenuation set as 10X. Only if the oscilloscope
fulfills the following two conditions at first, these specification standards can be
reached.
 This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
 If change of the operating temperature is up to or exceeds 5℃, do a
"Self-calibration" procedure (see "How to Implement Self-calibration" on
P12).
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Performance Characteristics
Bandwidth
Channel
Mode
Acquisition
Sample rate
(real time)
Input coupling
Input impedance
Input coupling
Max. input voltage
Channel –channel
Input
isolation
Time delay between
channel(typical)
Bandwidth limit
Sampling rate range
Interpolation
Max Record length
Scanning speed
(S/div)
Sampling rate / relay
Horizontal
time accuracy
System
Interval(△T) accuracy
(DC - 100MHz)
Vertical Resolution
Vertical
(A/D)
system
Sensitivity
8.Technical Specifications
100 MHz
4 channels
Normal, Peak detect, Averaging
1 GS/s
DC, AC , Ground
1 MΩ±2%, in parallel with 15 pF± 5 pF
1X,10X,100X,1000X
400V (DC+AC, PK - PK)
50Hz: 100 : 1
10MHz: 40 : 1
150ps
20 MHz, full bandwidth
0.5 S/s~1 GS/s
(Sinx)/x
20K
2 ns/div – 1000 s/div,
step by 1 – 2 - 5
± 100 ppm
Single:
± (1 interval time+100 ppm×reading+0.6
ns);
Average>16:
± (1 interval time +100 ppm×reading+0.4
ns)
8 bits (4 channels simultaneously)
5 mV/div~5 V/div