How to Set the Vertical System

The VERTICAL CONTROLS includes three menu buttons such as CH1, CH2,
CH3, CH4 and Math, and two knobs such as Vertical Position, Vertical
Scale.
Setting of CH1 – CH4
Each channel has an independent vertical menu and each item is set
respectively based on the channel.
To turn waveforms on or off (channel, math)
Pushing the CH1, CH2, CH3, CH4 or Math buttons have the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be
displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off
and its menu goes away.
The description of the Channel Menu is shown as the following list:
Function
Menu
Coupling
Inverted
Probe
MeasCurr
Limit
1. To set channel coupling
Taking the Channel 1 for example, the measured signal is a square wave
signal containing the direct current bias. The operation steps are shown as
4.Advanced User Guidebook
Setting
Description
Pass both AC and DC components of the input
DC
signal.
AC
Block the DC component of the input signal.
Ground
Disconnect the input signal.
ON
Display inverted waveform.
OFF
Display original waveform.
1X
10X
Match this to the probe attenuation factor to
100X
have an accurate reading of vertical scale.
1000X
OFF
Close meascurr
Turn the M knob to set the Amps/Volts ratio.
10A/V
The range is 100 mA/V - 1 KA/V.
100.0 mV/A
Amps/Volts ratio
Volts/Amp ratio is automatically calculated.
Full band
Get full bandwidth.
20M
Limit the channel bandwidth to 20MHz to
reduce display noise.
= 1/Resistor value