compensation is achieved (see Figure 3-6 and Figure 3-7).
Overcompensated
Figure 3-6 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if needed.

How to Set the Probe Attenuation Coefficient

The probe has several attenuation coefficients, which will influence the vertical
scale factor of the oscilloscope.
To change or check the probe attenuation coefficient in the menu of
oscilloscope:
(1) Push the function menu button of the used channels (CH1 - CH4 button).
(2) Select Probe in the right menu; turn the M knob to select the proper value
in the left menu corresponding to the probe.
This setting will be valid all the time before it is changed again.
Caution:
The default attenuation coefficient of the probe on the instrument is
preset to 10X.
Make sure that the set value of the attenuation switch in the probe is
the same as the menu selection of the probe attenuation coefficient in
the oscilloscope.
The set values of the probe switch are 1X and 10X (see Figure 3-8).
3.Quick Start
Compensated correctly
compensated
Figure 3-7 Adjust Probe