Proceed as follows to re-boot:
• Switch off the devices
◄
• Press the "left
" and "right
• Switch on the voltage again
• Then, the text "del
• Release the depressed buttons.
• Press the calibration button with a thin pencil or small screwdriver.
• Now the device is booted up and the display shows "Boot".
• Then, "done" appears in the display and the meter status of the main meter is dis-
played.
Then, re-transmit all device parameters to the TME400 or enter the values from the
test certificated.
Note
The serial interface is set to 38400 Bps, 8N1, Modbus RTU after booting.
These are also the default values of RMGView
(see chapter 4.4 RMGViewTME).
►
" buttons simultaneously
ll" appears in the display.
A
TME
Manual TME400-VMF · EN07 · December, 7
3 TME400
47
th
2020