Fig. 2-11
Inserting the eyepieces
Fig. 2-12
Inserting the eyepiece reticle
If eyepiece reticles are inserted into the unscrewed mount by the customer, attention must be
paid to the labelling being visible the right way up after insertion.
2-12
START-UP
Initial start-up
2.1.9
Insertion
centering telescope
• Remove both dust covers (2-11/1 and 4) from
the binocular tube.
• Remove both eyepieces (2-11/2) from the cases
and insert them in the binocular tube until stop.
• The centering telescope (2-11/3), which is used
to view the aperture/phase/darkfield stops and
to center phase and darkfield stops, can be
inserted in one of the tubes instead of an
eyepiece.
focusing on these stops and the setting to be
fixed using a clamping screw.
The eyepieces W-PL 10x/23 Br. foc.
and PL 10x/23 Br. foc. can be used
with
1-10).
Inserting the eyepiece reticle
The eyepieces PL 10x/23 Br. foc. are intended for
use with eyepiece reticles.
The slight image shift caused by the additional
path through the glass is taken into account on
the diopter scale by the fact that the zero point
position is indicated not by the white dot
(2-12/W), but by the red dot (2-12/R).
The eyepiece reticles (2-12/1) have been adhered
to screw-in mounts (2-12/2) by the manufacturer
to allow easy replacement.
To replace an eyepiece reticle, only remove screw-
in mount (2-12/2) with eyepiece reticle (2-12/1)
and replace it with a screw-in mount containing
the required eyepiece reticle.
Axioskop 2 FS plus
Axioskop 2 FS
of
eyepieces
The
adjustable
eyelens
all
tubes
(see
pp.
B 40-076 e