HBO 100 Microscope Illuminating System Operating Manual...
Page 3
Violations will entail an obligation to pay compensation. All rights reserved in the event of granting of patents or registration of a utility model. Published by: Carl Zeiss MicroImaging GmbH P.O.B. 4041, 37030 Göttingen, Germany Phone: +49 (0) 551 5060 660...
Please make yourself familiar with the contents of these operating instructions before attaching CAUTION! and starting up the HBO 100 Microscope Dangerous electrical voltage. Illuminating System. Additional information is available from our maintenance service or from CAUTION! authorized agencies.
Both the HBO 100 microscope illuminator and − The light of the HBO 103 short-arc lamps the HBO 103 lamp must be transported in their contains a high amount of UV light: original packaging.
Safety Microscope Illuminating System Carl Zeiss Maintenance, Care, Lamp breakage HBO 100 − Clean the housing of the microscope Hints on Maintenance and illuminator when it is dirty. Care To ensure perfect function of the microscope illuminating system, it is necessary to use only Do not use any solvents.
Instrument Description Microscope Illuminating System Carl Zeiss Components, Catalogue Numbers HBO 100 The HBO 100 Microscope Illuminating System Instrument is available in a manually and a self-adjusting Description version. The microscope illuminating systems include the following components (Fig. 1/Fig. 2): Components of the −...
Instrument Description HBO 100 Technical Data Carl Zeiss Technical Data of the Microscope Illuminating Systems HBO 100 illuminator Lamp housing Cat. No. 423010 HBO 100 illuminator, self-adjusting Cat. No. 423011-9901 Length 165 mm Width 166 mm Height 205 mm Connection to stand...
Instrument Description Microscope Illuminating System Carl Zeiss Technical Data, Line spectrum HBO 100 Fig. 3 Line spectrum of the HBO 100/103 mercury short-arc lamps B 46-0063 e 03/06...
Microscope Illuminating System Start-Up HBO 100 Lamp Insertion and Change Carl Zeiss Start-Up − Before inserting the lamp, make sure the Lamp Insertion and Change illuminating system has been disconnected from line power. − The lamps may only be changed when they are cooled down;...
Page 14
Start-Up Microscope Illuminating System Carl Zeiss Lamp Insertion and Change HBO 100 • Push clamping lever at lamp base (Fig. 4/4) Procedure: down and insert the new lamp into the • Switch off the transformer. mount in the bottom part of the illuminator •...
Do not look directly into the lamp or its System reflected light. Procedure: • Install the HBO 100 microscope illuminator on the microscope (see Operating Manual of microscope). • Connect the microscope illuminator to the transformer as described in the supplied operating instructions: Fig.
(Fig. 8/1) and horizontal adjustment (Fig. 8/2). • Deactivate the adjusting aid again. 1 Screw for vertical adjustment 2 Screw for horizontal adjustment 3 Collector control knob Fig. 8 Adjusting elements of HBO 100 for manual adjustment B 46-0063 e 03/06...
If the adjustment fails again, check whether the lamp has been inserted correctly in the 1 LED key base of the HBO 100 (pushed in as far as it will 2 Collector control knob Fig. 9 Adjustment of HBO 100, self-adjusting go).
Setting Offset Values HBO 100 On the self-adjusting illuminator, the position Setting Offset Values on of the lamp arc and its mirror image may be HBO 100 Microscope individually customized, if required, by the Illuminator, self-adjusting adjustment of offset values. (423011-9901) After the automatic adjustment (see Section 3.4), LED key (Fig.
Microscope Illuminating System Start-Up HBO 100 Inserting the Collector Attachment Carl Zeiss • Pull out clamping ring (Fig. 11/4). Inserting the Collector • Remove heat protection filter (Fig. 11/3). Attachment • Screw the collector attachment into the lamp as far as it will go (Fig. 11/1). The retaining ring on the collector attachment must point to the outside.