Connect The Instrument To The Line And Switch The Illumination On/Off - Zeiss Axioskop 2 FS plus Manual

Table of Contents

Advertisement

Carl Zeiss
2.1.12

Connect the instrument to the line and switch the illumination on/off

3 2
0 0
K
1 2
V
1 0
0 W
R E
M O
T E
1
2
Fig. 2-14
Connection of halogen illuminator
Fig. 2-15
Switch halogen illuminator on/off
via separate power unit
2-14
START-UP
Initial start-up
The Axioskop 2 FS plus / FS
not have their own line connection in order to
avoid voltage potentials in the microscope which
might affect electrophysiological examinations.
Accordingly, each of the HAL 100 and HBO 103
microscope illuminators are supplied by a separate
voltage supply unit.
Switch the HAL 100 halogen illuminator on or
off
The HAL 100 halogen illuminator is supplied via
the separate 12 V DC 100 W power unit and
directly switched on/off from there.
• Connect
12 V 100 W socket (2-14/1) of the separate
power unit (also see 2.1.13).
Should the halogen illuminator be equipped with a
3-pin plug (cable shielding), the third pin of the
plug can be removed without any risk of the
microscope being affected.
• Connect the power unit to the line via
connector (2-14/2) and cable.
The separate power unit can be connected to line
voltage between 100 and 240 VAC, 50 - 60 Hz.
The power unit sets itself automatically to the
available line voltage within this range.
• Switch HAL 100 halogen illuminator on/off via
line switch (2-15/1).
Axioskop 2 FS plus
Axioskop 2 FS
microscopes do
MOT
plug
of
halogen
B 40-076 e
MOT
illuminator
to
02/01