7.2 Cleaning of the calibration standard

If the calibration standard is soiled or covered by dust, it can be cleaned carefully
using window cleaner and a soft tissue.
A damaged or a polluted calibration standard may cause incorrect calibration
and therefore incorrect measuring results.

7.3 Calibration on calibration standard

If the instrument indicates that the calibration is outdated or if it is required by the
applicable regulations, the ZRP 6030+ has to be calibrated.
Ensure that the ZRP 6030+ is placed
correctly on the calibration standard. The
ZRP 6030+ will indicate incorrect placement
with an error message. The instrument has to
snap in audibly with a click sound.