Proceq ZRP 6030+ Instruction Manual page 18

Retroreflectometer
Table of Contents

Advertisement

When repeatedly experiencing a deviation greater than 15%, service or factory
calibration might be required. Please contact Proceq or your authorized
Proceq agent for further information.
Press the symbols
calibration function. Trigger the calibration by
pressing
window displays the measured values as well
as the deviation from the expected values in
percent. Successful calibration is indicated
through "Calibration OK". Confirm the
calibration by pressing
return to the measuring mode.
Calibration results with deviations greater
than 15 % are indicated with the message
"Calibration is not OK". This can be caused
by a soiled or damaged standard or a soiled
optic window. It is suggested to reject such a
calibration result by pressing
determining the reason for the deviation.
Clean the standard as described in chapter
15.2 "Cleaning" on page 44 and recalibrate.
The standard settings for R can be seen by
pressing
,
Remark:
You can change the calibration interval as
described in chapter 13.6.8 "Calibrate" on
page 40.