16 Status and error messages
16.1 Status message – "Battery voltage is critical"
16.2 Status message – "Please calibrate R"
16.3 Error message – "Calibration is not OK"
In order to prevent damage to the battery, the
instrument switches off automatically before
the battery is completely drained. The ZRP
6030+ announces this by displaying the
symbol
and a status message.
Charge the battery as described in chapter
14.3 "Charging" on page 43.
Press the warning triangle to get a more
detailed explanation.
To solve this, carry out the calibration as
described in chapter 7 "Calibrate" on page
16. If you are sure that it is not necessary to
carry out a calibration, you can press
"Acknowledge all" and the message will
disappear.
To change the calibration interval refer to
chapter 13.6.8 "Calibrate" on page 40.
Calibration results with deviations greater
than 15 % are indicated by "Calibration is not
OK". This can be caused by a soiled or
damaged standard or a soiled optic window.
It is suggested to reject such a calibration
result by pressing
the reason for the deviation. Clean the
standard as described in chapter 7.2
"Cleaning of the calibration standard" on
page 17 and recalibrate. When repeatedly
experiencing a deviation greater than 15%,
service or factory calibration might be
required. Please contact Proceq or your
Page 45
before determining