Calibrate

7.1 Calibration standard

Always protect the calibration standard from dust, moisture and other environmental
factors, keep it stored in the carrying case.
The calibration standard has been measured in the production and the calibration
value can be found on the label.
release
To remove the unit from the calibration standard, release it with one hand from
the snap lock and lift it with other hand.
After the expiry date of the calibration standard a factory calibration of the
ZRP 6030+ and its corresponding calibration standard is required. Contact
either Proceq or your authorized Proceq agent. The factory calibration is valid
for 2 years.
The calibration standard delivered with the ZRP 6030+ is not interchangeable
and is valid only for the instrument with the corresponding serial number.
snap lock
Page 16
Calibration standard
Label:
-Serial number
-Calibration value
- Validity