5.1 Performing a Calibration

A calibration check of the potential values can be done using a Calomel (HgCl) reference
electrode.
The principle consists in measuring the potential difference between the electrode to be
checked (the rod or the wheel electrode) and an electrode taken as a reference. The reference
electrode used is a saturated calomel electrode for laboratory purposes.
1. Fill the electrode to be tested with copper sulfate as described in this manual.
2. Connect the reference electrode to the Ground cable - Input (orange cable) of the
PM8500 Sensor Unit.
3. Connect either the rod or the wheel electrode to the Electrode - Input (blue cable).
4. Put both probes together to close the electrical circuit. Recommendation for wheel
electrodes: measure direct on the wooden spigot (copper sulphate side).