In the case of damage to the protective gasket, it should be replaced with the
new one included in the delivery kit.
Do the following:
- unscrew the four screws securing the holder and the protective grids with the
gasket in the lowering of the cover and remove them;
- replace the damaged gasket between the two grids;
- install the grids with a new gasket in place and fix them with the holder and
screws.

3.2 Verification of the radiometer

The radiometers should be verified after manufacture, repair or during use
(periodically, at least once a year).
3.2.1 Verification operations
During verification, the operations presented in Table 3.2 should be performed.
113