3.2.3 Verification conditions
Verification should be performed in compliance with the following conditions:
- ambient air temperature range within (20±5) °C;
- relative air humidity from 30 to 80 %;
- atmospheric pressure from 86 kPa to 106.7 kPa;
- natural background level of gamma radiation should not exceed 0.30 μSv/h;
- power supply voltage within (3.0± 0.2) V.
3.2.4 Verification procedure
3.2.4.1 External examination
During external examination, the radiometer should meet the following
requirements:
- the delivery kit should be completed as stated in item 1.3.1 of the present OM;
- labeling should be accurate;
- Quality Control Department seals should not be violated;