b) check the condition of the of the USB connector pins.
3.1.3.2 Delivery kit completeness check
Check if the delivery kit is complete according to
Table 1.2.
3.1.3.3 Operability check of the device
3.1.3.3.1 Operability check of the device and its procedure
are performed according to OM 2.3.3.
3.1.3.4 Battery status control
The device's battery status control is performed during
daily use, and before the long-term storage of the device. Follow
the next procedure:
- Switch on the device;
- Monitor the battery's state-of-charge according to the
indicator on the display;
- If the batteries are low, replace the batteries. If the device
was equipped with storage batteries, charge the batteries using
an external charger.

3.2 Verification of the device

Devices should be verified during operation (periodic
verification at least once a year) and after repair according to
verification methods, which are determined by regulations of the
central executive body, which ensures the formation of state
policy in the field of metrology and metrological activities, or
by national standards.
3.2.1 Presentation of verification results
3.2.1.1 Positive results of periodic or after-repair
verification are recorded in the table of Annex B or by issuing a
verification
certificate
measurement equipment.
3.2.1.2 If the device is acknowledged unfit for use after its
verification, it gets the certificate of inadequacy.
for
the
legislatively
46