ECOTEST STORA-ABG RKS-01 Operating Manual page 124

Radiometer-dosimeter
Table of Contents

Advertisement

Calculate the average value of surface alpha-particles flux density and main
relative error of measurement in accordance with the recommendations of DSTU
GOST 8.207:2008 standard.
The radiometer is acknowledged to have passed the verification if the relative
main error in percentage when measuring surface alpha-particles flux density does
not exceed 20+150/A, where A is a dimensionless value that is numerically equal to
the value of measured alpha-particles flux density in part./(cm
3.2.4.6 Presentation of verification results.
3.2.4.6.1 Satisfactory results of periodic verification and calibration after repair
are certified in the Table of Appendix E or by issuing a certificate of verification of
the legally regulated means of measuring equipment.
3.2.4.6.2. If, as a result of verification, the radiometer is found to be unsuitable
for use, it gets the Certificate of Inadequacy.
122