ECOTEST STORA-ABG RKS-01 Operating Manual page 20

Radiometer-dosimeter
Table of Contents

Advertisement

1.2.15 The radiometer remains operational after the impact of the following
external factors:
- high frequency sinusoidal vibrations (with a crossover frequency from 57 to
62 Hz) in the range of 10 to 55 Hz, 0.15 mm bias for frequency lower than the
crossover frequency;
- shocks with a shock pulse duration of 5 ms, total number of shocks (1000±10)
and maximum shock acceleration of 100 m/s
- shocks in shipping container with an acceleration of 98 m/s
duration of 16 ms (number of shocks - (1000±10) in each direction) or equivalent
shake tests;
- exposure in shipping container to ambient air temperature from - 25 to + 55
and relative humidity up to (95±3) % at + 35
- photon-ionizing radiation with exposure dose rate that corresponds to ambient
dose equivalent rate up to 10 Sv/h during 5 minutes.
2
;
о
С temperature;
18
2
, a shock pulse
о
С