Ion Genestudio S5 Instrument Run Times - Thermo Scientific Ion GeneStudio S5 User Manual

Hide thumbs Also See for Ion GeneStudio S5:
Table of Contents

Advertisement

Ion GeneStudio S5 instrument run times

Number of
Chip type
reads
Ion 510
Chip
2–3 M
4–6 M
Ion 520
Chip
3–4 M
15–20 M
Ion 530
Chip
9–12 M
Ion 540
Chip
60–80 M
Ion 550
Chip
100–130 M
Ion GeneStudio S5 Instrument User Guide
The instrument automatically restarts when the software update is complete.
Approximate Ion GeneStudio S5 instrument run times for a given chip type, read
length, and number of reads are shown in the following table. Run times indicate
typical time from run start to BAM file delivery (1 run/initialization).
Read
Ion GeneStudio S5
length/output
200 bases
0.3–0.5 Gb
400 bases
0.6–1 Gb
200 bases
0.6–1 Gb
400 bases
1.2–2 Gb
600 bases
0.5–1.5 Gb
200 bases
3–4 Gb
400 bases
6–8 Gb
600 bases
1.5–4.5 Gb
200 bases
10–15 Gb
200 bases
20–25 Gb
Ion GeneStudio S5 instrument run times
Instrument run time
Ion GeneStudio S5
Sequencer
Plus Sequencer
4.5 hours
10.5 hours
7.5 hours
3.5 hours
12 hours
5.5 hours
12 hours
5.5 hours
10.5 hours
21.5 hours
21 hours
19 hours
10 hours
11.5 hours
Chapter 3 Instrument operation
Ion GeneStudio S5
Prime Sequencer
3 hours
3 hours
5 hours
5 hours
3 hours
5.5 hours
5.5 hours
5 hours
4 hours
8 hours
6.5 hours
8 hours
7 hours
6.5 hours
8.5 hours
3
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

A38194A38195A38196