Alarms, Notifications, and Events

Alarm
Fatal alarms
Fluidics Calibration Needed.
Chip Cooler can not reach desired
temperature.
Drive sdX failed smartctl health
check.
Failed to mount the results
filesystem.
System sensor check failed ipmi -
sensors.
System sensor check failed, CC.CC V
(LL.LL/HH.HH)
CPU temperature check failed,
Physical Id X.
Ion GeneStudio S5 Instrument User Guide
If the Alarms/Notifications indicator appears, press the indicator to open the
Notifications screen, then press Alarms or Events see detailed messages.
1
1
Alarms and Notifications indicator.
Where X is the letter of the drive that
failed. Indicates that a drive is failing.
One or more drives are missing from
the RAID that makes up sw_results.
A component on the motherboard is
failing or has failed.
Where CC.CC = current voltage, LL.LL
= the low voltage threshold, HH.HH =
the high voltage threshold. Indicates
a voltage drop on the motherboard.
Where X is the processor number.
Indicates that the processor is over
heating.
Ion GeneStudio S5