Item
Pulse clean
Provides step by step instruction to perform a pulse clean
of the instrument if a solid‑state buffer leak occurs.
Upload
Allows upload of instrument diagnostics files for review
Diagnostics
by support personnel.
Ion GeneStudio S5 Instrument User Guide
Description
Appendix A Touchscreen reference
When/How to use
Use if directed to do so by Technical
Support as part of a troubleshooting
procedure:
1. Install consumables and a used
sequencing chip.
Note: Install Cleaning solution in
both the Wash solution (1) and
Cleaning solution (2) locations.
Use an empty Reagent cartridge.
2. Close the instrument door, then
press Next.
The pulse cleaning procedure starts
automatically.
For troubleshooting if directed to do so
by Technical Support.
Note: The system must be connected
to an active network port. If no network
is present, the file is pushed to the
Torrent Browser. You can then
download the file and email it to
Technical Support.
To retrieve the file using the Torrent
Browser, navigate to Help4Instrument
Diagnostics, then follow the on-screen
instructions.
A
Settings
27