Appendix A Touchscreen reference
Settings
Perform the
instrument Clean
operation
32
The Clean program is normally performed automatically at the completion of the
previous sequencing run. Perform a Clean if the sequencing run:
• was aborted or had a power failure during the second of two runs started after a
single initialization.
• is not completed and >200 flows occurred before the stoppage.
• the post-sequencing run cleaning was not completed.
IMPORTANT! Do NOT perform a manual cleaning with an unused, initialized Ion
S5
Sequencing Reagents cartridge. Perform an instrument reset run instead. See
"Perform an instrument reset run with an initialized, unused Sequencing Reagents
cartridge" on page 18 for more information.
1. In the Settings menu, press Clean Instrument.
The Clean procedure begins automatically.
2. Follow the on-screen instructions (see page 16 for more information), then press
Next.