Thermo Scientific Ion GeneStudio S5 User Manual page 40

Hide thumbs Also See for Ion GeneStudio S5:
Table of Contents

Advertisement

A
Appendix A Touchscreen reference
Alarms, Notifications, and Events
Alarm
Front/Rear FanX speed below
threshold YYYY.
FPGA failed to connect in emulate
mode.
Valve Link Failure.
OIA (On Instrument Analysis) is not
running.
FPGA temperature is above the
threshold of 70°C.
Boot drive exceeded 75%.
RAID degraded one disk has failed.
RAID failure detected, more than one
disk failed.
/sw_results directory not mounted.
No GPU detected.
No GPU Driver detected.
Air Compressor failure.
Non fatal alarms
Instrument must be cleaned before
starting another run.
Ambient temperature is above the
threshold of 35°C.
RAID degraded one disk has failed.
No Connectivity to Torrent Server.
No Connectivity to FTP server.
40
Description
Where X is the fan number and YYYY
is the low threshold speed for the
system fan.
Internal error.
One or more of the system fans has
failed.
The boot drive is unexpectedly full.
One of the Torrent Server drives has
failed. System is still operational but
must be serviced soon.
More than one of the Torrent Server
drives has failed. System is no longer
operational.
One or more of the drives used during
data collection has failed.
GPU has failed.
Air compressor is either leaking or
has stopped working.
Maximum number of runs since last
clean has been exceeded.
Verify the temperature in the lab.
One of the Torrent Server drives has
failed. System is still operational but
must be serviced soon.
Check that the Ethernet cable
connecting the instrument to the
server is properly connected, see
page 13 for the Ethernet port
location.
Recommended Action
Contact Technical Support.
Perform an instrument cleaning. See
"Perform the sequencer cleaning
manually" on page 16.
Contact Technical Support if the
temperature of the lab is within
normal operating temperature (20°C
to 30°C).
Contact Technical Support.
Reconfigure the Torrent Server as
needed, see "Network Settings" on
page 20.
Reconfigure the FTP server as
needed, see "Network Settings" on
page 20.
Ion GeneStudio S5 Instrument User Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

A38194A38195A38196

Table of Contents