Specifications - QuadTech Guardian 2500 Series Instruction Manual

Ac/dc/ir hipot tester
Table of Contents

Advertisement

Guardian 2510/2520/2530
AC Output Voltage:
Voltage Display:
AC Current Display: Real, Imaginary or Total current (user selectable)
Guardian 2520/2530
DC Output Voltage:
Voltage Display:
DC Current Display: Range:
Guardian 2530
Insulation Resistance: Voltage:
Charging Current:
Voltage
1000V
500V
50V
50V
10K

Specifications

Range:
0.1 to 5KV AC, in 2V steps
Regulation:
+/-(1% of setting +5V)
Frequency:
50/60Hz selectable, 100ppm accuracy
Waveform:
Sinusoidal, crest factor 1.4
Accuracy:
+/-(1% of reading +5V)
Resolution:
1V steps
Range:
1µA to 15mA AC, in 1µA steps
Accuracy:
+/-(0.5% of set high limit + 1µA) *
Range:
0.1 to 6KV DC, in 2V steps
Accuracy:
+/-(1% of reading +5V)
Resolution:
1V steps
0.1µA to 1.0mA DC, in 0.1µA steps
1.0mA to 7.5mA DC, in 1µA steps
Accuracy:
+/-(0.5% of set high limit + 0.1µA or 1µA) *
50 - 1000V DC in 1V steps
Accuracy:
+/-(2% of setting + 5V)
Range:
10kΩ - 2TΩ, voltage dependent
(range limited to V/50µA in DUT Grounded mode)
Accuracy:
+/-(2% + 1 count) for V/R ≥ 10nA
+/-(5% + 1 count) for V/R < 10nA
(
Accuracy above V/10nA is degraded with
Ground Continuity (GC) function in use)
7.5mA maximum
2% Accuracy
150K
*Accuracy is degraded by a factor of 2 in DUT
Grounded mode
5% Accuracy
5G
100G
Resistance
2T
Page 9 of 9