Optical laser measurement system
Table of Contents

Advertisement

Display and control panel AMS 301i
I/O submenu
Table 8.4:
Level 3
I/O 1
I/O 2
Limit values
43
I/O submenu
Level 4
Level 5
Selection/configuration option
Description
Port config-
Input/Output
uration
Defines whether I/O 1 functions as an output or input.
Switching
Function
No function/preset teach/laser ON/OFF
input
Activation
Low active/High active
Switching
Function
Pos. limit value 1 / Pos. limit value 2 / Velocity / Intensity (ATT) /
output
Temp. (TMP) / Laser (LSR) / Plausibility (PLB) / Hardware (ERR)
The individual functions are "ORed" on the selected switching
output.
Activation
Low active/High active
Port config-
Input/Output
uration
Defines whether I/O 2 functions as an output or input.
Switching
Function
No function/preset teach/laser ON/OFF
input
Activation
Low active/High active
Switching
Function
Pos. limit value 1 / Pos. limit value 2 / Velocity / Intensity (ATT) /
output
Temp. (TMP) / Laser (LSR) / Plausibility (PLB) / Hardware (ERR)
The individual functions are "ORed" on the selected switching
output.
Activation
Low active/High active
Upper pos.
Activation
ON / OFF
limit 1
Limit value
Value input in mm or inch/100
input
Lower pos.
Activation
ON / OFF
limit 1
Limit value
Value input in mm or inch/100
input
Upper pos.
Activation
ON / OFF
limit 2
Limit value
Value input in mm or inch/100
input
Lower pos.
Activation
ON / OFF
limit 2
Limit value
Value input in mm or inch/100
input
i
AMS 301
Standard
Output
No function
Low active
Plausibility (PLB),
hardware (ERR)
Low active
Output
No function
Low active
Intensity (ATT),
Temp. (TMP),
Laser (LSR)
Low active
OFF
0
OFF
0
OFF
0
OFF
0