Page 3
307i The main menus Device information - main menu AMS 307i 120 Leuze electronic This menu item contains detailed information on • Device model, GmbH & Co. KG • Manufacturer, SW: V 1.3.0 HW:1 • Software and hardware version, SN: ----------------------- •...
Notice! The Declaration of Conformity for these devices can be requested from the manufacturer. The manufacturer of the product, Leuze electronic GmbH + Co. KG in D-73277 Owen, possesses a certified quality assurance system in accordance with ISO 9001. LISTED...
With the AMS 3xxi product series, Leuze electronic makes available a range of internation- ally relevant interfaces. Note that each interface version listed below corresponds to a different AMS 3xxi model.
Leuze electronic GmbH + Co. KG is not liable for damages caused by improper use. Read the technical description before commissioning the device. Knowledge of this technical description is an element of proper use.
The device must not be tampered with and must not be changed in any way. The device must not be opened. There are no user-serviceable parts inside. Repairs must only be performed by Leuze electronic GmbH + Co. KG. Competent persons Connection, mounting, commissioning and adjustment of the device must only be carried out by competent persons.
Safety Exemption of liability Leuze electronic GmbH + Co. KG is not liable in the following cases: • The device is not being used properly. • Reasonably foreseeable misuse is not taken into account. • Mounting and electrical connection are not properly performed.
Page 12
Affix the laser information and warning signs so that they are legible without exposing the reader to the laser radiation of the device or other optical radiation. A Laser aperture B Laser warning sign C Laser information sign with laser parameters Figure 2.1: Laser apertures, laser warning signs Leuze electronic 307i...
Further information can be found in chapter 5.2 and chapter 5.3. 3.1.2 Mounting the reflector The reflector is mounted using 4 screws (M5). The reflector is angled using the spacer sleeves included. Incline the reflector by approx. 1°. Detailed information can be found in chapter 6.4. Leuze electronic 307i...
• Number of data bits • Resolution • Error bit on/off For further information, see also chapter 9 "SSI interface" on page 52 and display menu structure level 2, menu item "SSI" in the appendix of the manual. 307i Leuze electronic...
655 nm ≤ 18μs Impulse duration ≤ 4mW Max. output power (peak) Interfaces SSI clock rate 50kHz … 800kHz Operating and display elements Keyboard 4 buttons Display monochromatic graphical display, 128 x 64 pixels 2 LEDs, two-colored Leuze electronic 307i...
Page 17
This is a Class A product. In a domestic environment this product may cause radio inter- ference, in which case the operator may be required to take adequate measures. The AMS 307i is designed in accordance with safety class III for supply with PELV (protective extra-low voltage). 307i Leuze electronic...
Technical data 4.1.2 Dimensioned drawing AMS 307i M5 screw for alignment Knurled nut with WAF4 hexagon socket and M5 nut for securing Optical axis Zero point of the distance to be measured Figure 4.1: Dimensioned drawing AMS 307 Leuze electronic 307i...
Device name plate using the AMS 358i as an example Notice! Please note that the shown name plate is for illustration purposes only; the contents do not correspond to the original. Save the original packaging for later storage or shipping. Leuze electronic 307i...
Installation and mounting If you have any questions concerning your shipment, please contact your supplier or your local Leuze electronic sales office. Observe the applicable local regulations when disposing of the packaging materials. Mounting the AMS 307i M5 screw for align-...
Page 22
WAF4 hexagon socket ("B" in figure 5.2). Knurled nut and nut must not be tightened until alignment has been completed. Attention! The device must not be opened. Failure to comply will render the guarantee void. Warranted features cannot be guaranteed after the device has been opened. Leuze electronic 307i...
Optional mounting bracket A mounting bracket for mounting the AMS 307i on a flat, horizontal surface is available as an optional accessory. Type designation: MW OMS/AMS 01 Part no.: 50107255 Laser beam Figure 5.3: Optional mounting bracket 307i Leuze electronic...
X = 100mm + (max. measurement distance in mm x 0.01) The AMS 307i are mounted stationary and in parallel on one plane. Both reflectors move in parallel at the same distance to the AMS 307i. Measurement distance up to 120m: minimum parallel spacing X ≥ 600mm Leuze electronic 307i...
Depending on the size of the used reflector, the DDLS can be mounted with a minimum parallel spacing of 100mm to the AMS 307i. The parallel spacing is independent of the distance. 307i Leuze electronic...
Use the M5 screws to mount the deflector unit. Secure the screws with a toothed lock washer to protect against loosening caused by vibrations. Figure 5.5: Mounting variants of the US AMS 01 laser beam deflector unit Leuze electronic 307i...
Installation and mounting 5.3.2 Dimensioned drawing of US AMS 01 deflector unit Laser beam Figure 5.6: Dimensioned drawing of US AMS 01 deflector unit 307i Leuze electronic...
Height of the spring in non- preloaded state A Mirror Figure 5.7: Photo and dimensioned drawing of the US 1 OMS deflector unit Alignment of the laser light spot on the reflector is performed as described in chapter 5.2. Leuze electronic 307i...
All provided specifications for the AMS 307i, such as the operating range or accuracy, can only be achieved with the reflective tape specified by Leuze electronic. The reflective tapes are available as adhesive tapes, affixed to a metal plate and with an integrated heater especially for use at low temperatures.
Do not use any agents that act with a grinding effect. A conventional household Cleaning detergent can be used as a cleaning agent. Rinse with clear water and dry the surface. Storing the reflector Store in a cool and dry place. Leuze electronic 307i...
Dimensioned drawing of reflective tape on a support plate Always align the TOP marking with the AMS connections! (chapter 6.4.2) Figure 6.1: Dimensioned drawing of reflectors Part Reflective tape (mm) Reflector plate (mm) Reflective tape 200x200-M Reflective tape 500x500-M Reflective tape 914x914-M 307i Leuze electronic...
Do not use any agents that act with a grinding effect. A conventional house- Cleaning hold detergent can be used as a cleaning agent. Rinse with clear water and dry the surface. Storing the reflector Store in a cool and dry place. Leuze electronic 307i...
Attention! The reflector sizes shown below are a recommendation from Leuze electronic for on-vehicle mounting of the AMS 307i. For stationary mounting of the AMS 307i, a smaller reflector is generally sufficient for all measurement distances.
M12 connections are on the top, the "TOP" label of the reflector is also on the top. If the AMS 307i is mounted so that the M12 connections are on the side, the "TOP" label of the reflector is also on the side. 307i Leuze electronic...
Page 36
Reflective tapes …-S and …-M Pitch approx. 1° Direct reflection from the triad structure Deflected surface reflection due to the pitch of the reflective tape Figure 6.3: Mounting the reflector Pitch approx. 1° Spacer sleeves Figure 6.4: Pitch of the reflector Leuze electronic 307i...
Page 37
Pitch approx. 1° Direct reflection from the triad structure Deflected surface reflection due to the pitch of the reflective tape Figure 6.5: Mounting of heated reflectors Pitch approx. 1° Spacer sleeves Figure 6.6: Pitch of the heated reflector 307i Leuze electronic...
Notice! Reliable function of the AMS 307i and, thus, max. operating range and accuracy can only be achieved with the reflective tape specified by Leuze electronic. No function can be guar- anteed if other reflectors are used! Leuze electronic 307i...
For UL applications, use is permitted exclusively in Class 2 circuits according to NEC (Na- tional Electric Code). The laser measurement systems are designed in accordance with safety class III for supply by PELV (protective extra-low voltage with reliable disconnection). 307i Leuze electronic...
Transmission line RS 232/service data Voltage supply 0VDC RS232-RX Receiving line RS 232/service data Not used RS232-RX M12 socket Thread Functional earth (housing) (A-coded) Table 7.3: Service pin assignment Notice! The service interface is designed only for use by Leuze electronic! 307i Leuze electronic...
Input 2 or output 2 active: Function depending on configuration. Warning - laser prefailure message: Laser diode old, device still functional, exchange or have repaired. Warning - temperature monitoring: Permissible internal device temperature exceeded / not met. Leuze electronic 307i...
Page 43
The measured position value is displayed in the configured unit of measurement. With the metric setting, the measurement value is always displayed in meters +87.000m with three decimal places. +87.0in With the inch setting, the measurement value is always displayed in inches with one decimal place. 307i Leuze electronic...
- No measurement value output; for details, see Display Orange continuous light Power LED orange - Parameter enable active - No data on the host interface BUS LED BUS LED off - No supply voltage (Power) - SSI interface deactivated Leuze electronic 307i...
Default ----------------- unit Save Use the buttons to set the desired value. An accidental, incorrect entry can be corrected by selecting <-| and then pressing Then use the buttons to select Save and save the set value by pressing 307i Leuze electronic...
The display then shows the measurement window with all status information. Device information - main menu AMS 307i 120 This menu item contains detailed information on Leuze electronic • Device model, GmbH & Co. KG • Manufacturer, SW: V 1.3.0 HW:1 •...
Parameters to default, all parameters are reset to their stan- dard settings without any further security prompts. In this case, English is selected as the display language. Additional important information on parameter handling can be found at the end of the chapter. 307i Leuze electronic...
Page 48
The preset value is accepted by means of teach pulse. The teach pulse can be applied to a hardware input of the M12 PWR connector. The hard- ware input must be appropriately configured. See also configuration of the I/Os. Leuze electronic 307i...
Page 49
ON / OFF limit 1 Limit value Value input in mm or inch/100 input Upper pos. Activation ON / OFF limit 2 Limit value Value input in mm or inch/100 input Lower pos. Activation ON / OFF limit 2 307i Leuze electronic...
Page 50
Service RS232 Baud rate 57.6kbit/s / 115.2kbit/s 115.2kbit/s The service interface is only available to Leuze internally. Format 8,e,1 / 8,n,1 8,n,1 The service interface is only available to Leuze internally. Leuze electronic 307i...
The display language is a passive operational control and is, thus, not a func- tion parameter, per se. 8.3.4 Service menu Service Status messages Diagnostics Expanded diagnostics A more detailed description of the individual functions can be found in chapter 10. 307i Leuze electronic...
Viewing and editing parameters As long as parameter enabling is activated, the entire AMS 307i display is inverted. Notice! If a password was stored, parameter enabling is not possible until this password is entered, see "Password for parameter enabling" below. Leuze electronic 307i...
Page 53
Press the enter button to enter the Password menu. Activate password In the Activate password menu, use the buttons to select the ON menu item and press the enter button Press the ESC button to return to the Password menu. 307i Leuze electronic...
Page 54
See "Setting values" on page 42. <-|0123456789 save Default ----------------- unit 0000 Press the ESC button twice to return to the Param- eter menu. Notice! The master password 2301 can enable the AMS 307i at any time. Leuze electronic 307i...
The SSI interface can only represent positive distance values. If negative output values are ascertained due to the offset or count direction, a zero value is output at the SSI interface! In the event of a number overflow, all data bits are set to "1". 307i Leuze electronic...
The measurement value on the SSI interface of the AMS 307i is updated every 1.7ms (de- fault) independent of the clock frequency. The update rate on the interface can be reduced to 0.2ms via the display under the SSI menu item. Leuze electronic 307i...
For basic operation of the display please refer to chapter 8.2.3. In order to change the parameters please activate parameter enabling. The SSI interface remains active even during parameter enabling. Changes to parameters have an immediate effect. Leuze electronic 307i...
(always "1", since no summation occurs) The status messages within the ring memory are selected with the up/down buttons The enter button can be used to call up detailed information on the corresponding status messages with the following details: 307i Leuze electronic...
Leuze for internal evaluation. The diagnostics have no influence on the communication to the host interface and can be activated during operation of the AMS 307i. 10.1.3 Expanded diagnostics The Expanded diagnostics menu item is used for Leuze-internal evaluation. Leuze electronic 307i...
Interface errors 10.3.1 BUS LED Error Possible error cause Measure No supply voltage connected Check supply voltage. BUS LED "OFF" Incorrect wiring Check wiring. SSI deactivated Activate SSI interface in the AMS 307 . Table 10.2: Bus error 307i Leuze electronic...
DIN 47100 Specifications of the SSI connection cable Operating temperature range in rest state: -40°C … +80°C in motion: -5°C … +80°C Material free of halogens, silicone and PVC Bending radius > 80mm, suitable for drag chains Leuze electronic 307i...
Page 67
20m KB SSI/IBS-25000-BA M12 socket, B-coded, for SSI/Interbus, axial connector, 50108447 open cable end, cable length 25m KB SSI/IBS-30000-BA M12 socket, B-coded, for SSI/Interbus, axial connector, 50108446 open cable end, cable length 30m 307i Leuze electronic...
Contact your Leuze distributor or service organization should repairs be required. The addresses can be found on the inside of the cover and on the back. Notice! When sending the laser measurement systems to Leuze electronic for repair, please provide an accurate description of the error. 12.3...
Page 69
Overview of reflector types ....60 Fast commissioning ....11 307i Leuze electronic...
Page 71
Menu structure AMS 307i Level 1 Level 2 Level 3 Level 4 Level 5 Selection/configuration option Detailed infor- mation on : selection : selection : selection : selection : selection : selection : back : back : back : back : activate : back Device information...
Page 72
Deutsch / English / Español / Français / Italiano page 48 Service Status messages Number of readings, reading gates, reading rate / non-reading rate etc. page 48 Diagnostics Exclusively for service purposes by Leuze electronic Expanded diagnostics Exclusively for service purposes by Leuze electronic...