Standard Calibration Sensor Counts And Levels Record; Calibration Procedures; Setting The Low Level And Collecting Cal Low Data - Vega FiberFlex LFXG-H Operating Instructions Manual

Detector with gen2000 electronics for level measurement
Table of Contents

Advertisement

Calibration procedures

Standard method
Includes these steps (see the following pages):
1.

Setting the low level and collecting Cal Low data*

2.
Setting the high level and collecting Cal High data*
3.
Collecting the linearizer table data*
4.
Calculating the linearity
5.
Calculating the calibration
* Perform these data collection steps in any sequence, depending on your
ability to empty and fill the vessel.
If using the standard method, it may be helpful to record the sensor counts
and levels at each step:

Standard calibration sensor counts and levels record

Table 3.2
Data type
Cal low level (usually empty)
Linearizer data point
Cal high level (usually full)
Linearizer data point
1 Setting the low level and collecting Cal Low data
You must:
1. Use the gauge to measure the low process level.
LFXG-H Installation and Operation Guide
Sensor counts
0
1
2
3
4
5
6
7
8
9
10
Simple method
Skips steps 3 and 4.
Actual level
(engineering units)
Calibration
51

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents