Table 4.2
Process Variables tab — display values
Value
Source Decay Gain
Stdz Gain
HV Setting

Gauge Info tab

Gauge Info tab — additional display values
Table 4.3
Value
Sensor Coefficients
LFXG-H Installation and Operation Guide
Description
The current value of the source decay gain. Use this to
compensate for the natural decay of the radiation source,
which produces a lower field over time.
The current value of the standardize gain that adjusts with
each standardize procedure.
The set point for the sensor high voltage.
Description
The algorithm that compensates for variations in
measurement output with changes in temperature
uses temperature coefficients. The factory
determines the coefficients through rigorous
testing. You cannot change them through normal
operation.
Advanced functions
Scintillator sensor voltage
Firmware version on the
FLASH
Hardware's version number
GEN2000 CPU board's
serial number
GEN2000 unit's serial
number
Sensor Coefficients
T0 – T3