Jumpers - Vega FiberFlex LFXG-H Operating Instructions Manual

Detector with gen2000 electronics for level measurement
Table of Contents

Advertisement

Table 5.5
CPU test point labels
Label
Description
Count
Raw input signal coming from the preamplifier.
GND
Logic ground
U5 pin 8
+5 V power supply test points. Referenced to logic ground.

Jumpers

Jumpers JP1 and JP2 on the power supply board set the current loop
source or sink mode.
Note: Do not change the jumpers from the current setting without
Jumper settings
Table 5.6
Mode
Source mode
Sink mode
The level gauge does not use jumpers J1 – J4 on the CPU board.
LFXG-H Installation and Operation Guide
calling VEGA Field Service.
Gauge current loop
Self-powered
DCS-powered
Diagnostics and repair
Jumper setting
JP1 1-2, JP2 2-3
JP1 2-3, JP2 1-2
85