Correction Factor; Table 5:3 - Correction Factors - Xylem SonTek FlowTracker2 User Manual

Table of Contents

Advertisement

a) For each measurement, you must manually enter the measurement depth.
b) The mean station velocity (mean velocity in vertical) is calculated by integrating all
velocity measurements considering their location as illustrated in
c) If multiple measurements are made at the same measurement depth, these
measurements are averaged prior to calculating the velocity integration.
5.1.3.3

Correction Factor

The Correction Factor (CF) is a user supplied parameter used to scale the mean
station velocity or velocity in the vertical.
Be careful when selecting a Correction Factor, as the mean station velocity is always
scaled by this value. Incorrect use of this parameter will affect the final discharge
measurement.
a) The CF is commonly used at edges, internal islands, and other method None
stations to scale the velocity from the adjacent station(s).
b)
Table 5:3
list the Correction Factors implemented for each Station Type and
Velocity Method.
Station Type
Bank
Island Edges
Open water
FlowTracker2 User's Manual (January 2019)
Figure 5:5 - Vertical Velocity Curve
Table 5-3 – Correction Factors
Velocity Method
n/a
n/a
None
0.6
78
SonTek – a Xylem brand
Figure 5:5.
Correction Factor
User defined
User defined
User defined
1

Advertisement

Table of Contents
loading