Table 14:2. Original Flowtracker File Format Import - Xylem SonTek FlowTracker2 User Manual

Table of Contents

Advertisement

Figure 14:24. Original FlowTracker .wad File Import Tool
The FlowTracker2 software supply two options to import Original FlowTracker data from
either the ASCII or WAD files, shown in
The criteria associated with each file format import are summarized in
of ASCII files will preserve the Original FlowTracker discharge calculations. If any
editing is performed in FlowTracker2 software, the Original FlowTracker discharge will be
recalculated based on FlowTracker2 Engine.
Table 14-2. Original FlowTracker File Format Import
File
Calculation
Format
ASCII
Not Recalculated
WAD
Recalculated
ASCII: The "Put Headers on ASCII Files" option in the Original FlowTracker desktop
software must be selected before the ASCII files are created. This will ensure that
header information is placed at the top of the ASCII file.
FlowTracker2 User's Manual (January 2019)
Figure
14:24.
Figure 14:25. FlowTarcker1 Import
Engine
Original FlowTracker
Engine
FlowTracker2 Engine
262
SonTek – a Xylem brand
Table 14:2
Requirement
The (4) ASCII and WAD files
must be present in the same
folder.
The .DIS, .DAT, .CTL, and
.SUM ASCII files are exported
using the Original FlowTracker
desktop software
The WAD file must be present
in the folder.
.