Lab-Adv Configuration; Table 15:1. Configuration Settings - Xylem SonTek FlowTracker2 User Manual

Table of Contents

Advertisement

15.3. Lab-ADV Configuration

The Lab-ADV Configuration settings specified in
data collection process within Communication, System, Environmental and Data
Management categories. The configuration settings are application dependent and will
vary depending on the specific measurement requirements.
Category
Communication COM Port
Sampling
FlowTracker2 User's Manual (January 2019)
Figure 15:2 – Upload of License File for Lab-ADV
Table 15-1. Configuration Settings
Value
Sampling Rate
Velocity Range
Table 15:1
allows the user to define the
Description
COM port assigned to FlowTracker2 probe
assembly. It is recommended that the
Windows Device Manager be used to
determine assigned COM port.
Sampling rate selected for measurement.
The Lab-ADV has four sampling rates: 1,
2, 5, and 10 Hz.
Optimizes ADV operation for expected
maximum velocity conditions. There are
six velocity range settings: +/- 10, 20, 50,
100, 200 cm/s and Auto. Auto velocity
range is the most robust setting under
variable or unknown conditions. Auto is
not available at the10 Hz sample rate.
287
SonTek – a Xylem brand

Advertisement

Table of Contents
loading