9.3.2.2

Review Point Measurement

Review Point Measurement is a report supplied at the end of each point velocity
measurement. The report consists of two components, tabular summary of numerical
calculations performed on the raw data collected and graphical display of raw data.
Review Point Measurement screen consists of the
following tabular summary and graphical displays,
Tabular Display
a) Review Point Measurement,
Graphical Display of raw data are based on the
following variables,
b) Velocity,
c) SNR,
d) Temperature,
e) Tilt,
f) Velocity Angle,
g) Battery.
h) To select a variable,
i). Use the up or down arrow key to view a
variable,
i) To accept the velocity measurement,
i). Press the Right Soft Key.
j) To
REDO
velocity measurement,
i). Press the Left Soft Key.
The conventions and graphic display features used for displaying Velocity and SNR raw
data are defined under
Properties associated with the variable output from the numerical calculations
performed on raw data collected is defined in
Variable
Warning supplied based on quality control
Warning
criteria
Samples
Total number of sample received
Spikes
Number of spikes removed from mean velocity
Mean X velocity component of all samples.
Vel. X
Velocity is despiked and mounting correction
applied (if applicable).
Mean Y velocity component of all samples.
Vel. Y
Velocity is despiked and mounting correction
applied (if applicable).
FlowTracker2 User's Manual (January 2019)
Table 9:4
Raw Data
Display.
Table 9-4 - Review Point Measurement
Description
164
SonTek – a Xylem brand
Figure 9:15 - Review Point
Table
9:4,
Measurement