Step 4
Fractional Depth Measurement screen displays
graphical display of raw data of the following variables
during a velocity measurement,
a) Velocity,
b) SNR,
c) Temperature,
d) Tilt,
e) Vel. Angle,
f) Battery.
Tilt and Velocity Angle indicators are situated on the
right hand side of the screen.
g) Tilt,
h) Velocity Angle,
i) Averaging Time left.
j) To select a variable graphical display,
i). Use up or down scroll arrows keys to view a
variable.
k) To stop sampling,
i). Press the Right Soft Key.
l) To
CANCEL
velocity measurement,
i). Press the Left Soft Key.
The conventions and graphic display features used for displaying Velocity and SNR raw
data are defined under
8.3.3.3

Review Point Measurement

Review Point Measurement is a report supplied at the end of each point velocity
measurement. The report consists of two components, tabular summary of numerical
calculations performed on the raw data collected and graphical display of raw data.
Review Point Measurement screen consists of the
following tabular summary and graphical displays,
Tabular Display
a) Review Point Measurement,
Graphical Display of raw data are based on the
following variables,
b) Velocity,
c) SNR,
d) Temperature,
e) Tilt,
f) Velocity Angle,
FlowTracker2 User's Manual (January 2019)
Raw Data
Display.
Table 8:8