Table of Contents

Advertisement

Edit
FT
Orig.
• Automatic beam check – For each station, the mean beam check data is
displayed below the station information summary table. The SNR for each ADV
beam is plotted with distance ("Range") from the sample volume.
14.8.2.3

Samples

The Samples tab
(Figure
selection can be performed by selecting the appropriate station number in the "Station
Number" column on the left of plots. All point measurements at the same station will be
allocated the same station number. The header of each plot distinguishes between the
specific station number and point measurement.
• Velocities – X,Y, (and Z, if using a 3D probe) velocities measured during a station
point measurement spanning chosen sampling period,
• SNR – beam SNR measured during a station point measurement spanning
chosen sampling period,
• Temperature – Temperature measured during a station point measurement
spanning chosen sampling period,
FlowTracker2 User's Manual (January 2019)
FT2
Value
satellites
Fix quality
HDOP
Figure 14:28. Layout of Samples
14:28) contains the following plots. The point measurement
Description
was recorded
GPS fix quality if GPS fix was
recorded
HDOP if GPS fix was recorded
268
SonTek – a Xylem brand