Maintenance Task Schedule; How Often Should Maintenance Tasks Be Performed - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Care and Maintenance

10-4 Maintenance task schedule

10-4-1 How often should maintenance tasks be performed?

NOTE:
NOTE:
Service at Indicated
Time
Clean Probes
Inspect AC Mains Cable
Inspect Cables and
Connectors
Clean Console
Clean LCD
10-6
The Care and Maintenance task schedule (provided in
Table 10-1 on page
T8 Pro should be serviced and outlines items requiring special
attention.
It is the customer's responsibility to ensure the Vivid T8/Vivid T8
Pro care and maintenance is performed as scheduled in order to
retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowledge of
your Vivid T8/Vivid T8 Pro and can best provide competent,
efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in
the Care and Maintenance Task Schedule assumes that you
use your Vivid T8/Vivid T8 Pro for an average patient load
(10-12 per day) and not use it as a primary mobile Ultrasound
system which is transported between diagnostic facilities.
If conditions exist which exceed typical usage and patient load,
then it is recommended to increase the periodic maintenance
frequencies.
Table 10-1:
Customer Care Schedule
Daily
Weekly
•*
Vivid T8/Vivid T8 Pro – Basic Service Manual
10-6) specifies how often your Vivid T8/Vivid
Per
Facilities
QA
Monthly
Program
5718826-100 English Rev. 10
Notes
* or before each use
Mobile Ultrasound
system: Check
Weekly